OX60

Portable digital dissolved oxygen meter


WPA’s OX60 meter is based on proprietary Application Specific Integrated Circuit (ASIC) design and uses a new low cost 12mm diameter galvanic DO probe that is easy to use. This compact meter is ideal for indoor and outdoor applications.
The OX60 Dissolved Oxygen meter allows for customisation by the discerning user. Functions include: Auto-off; % saturation; mg/l or ppm mode; temperature compensation; barometric pressure & salinity correction; electrode diagnosis; user reset and restore to factory defaults.
At the push of a button, the meter switches measurement mode from mg/l or ppm to % saturation.
100% or 0% saturation calibration is easy at the press of a button. mg/l or ppm calibration can also be performed using a known oxygen concentration solution or reference DO meter.

The meter is supplied with DO Electrode, membrane cap, refilling solution and carrying case

DOCUMENTS


FEATURES

  • Protective Rubber Boot
    shields meter from accidental knocks and features a sturdy built-in stand for easy bench-top operation
  • User Customisation
    for user's flexibility like toggling of various modes from ppm, mg/l to % saturation, and viewing electrode diagnosis
  • Easy Calibration
    can be performed at 100% and/or 0% and mg/l or ppm against a known solution
  • Automatic Temperature Compensation (ATC)
    guarantees the highest accuracy under varying temperature condition
  • Galvanic Cell
    allows instantaneous measurements, producing quick responses and eliminates polarisation; requires minimal maintenance
  • Automatically Compensates Salinity and Barometric Pressure
    after manual input for increased accuracy
  • Hold Function
    momentarily freezes reading for easy viewing
  • Auto-Power Off
    saves battery power after non-use
  • Large Custom LCD
    provides optimum viewing even at a distance
  • Self-Diagnostic
    with message codes for easy troubleshooting

SPECIFICATION

Dissolved Oxygen Range 0 to 20.00 mg/l or ppm
Resolution & Accuracy 0.01 mg/l or 0.01 ppm & ±1.5% Full Scale % Saturation of Oxygen 0.0 to 200/0 %
Resolution & Accuracy 0.1% & ±1.5% Full Scale
Temperature Range 0.0 to 100.0 °C
Resolution & Accuracy 0.1 °C & ± 0.3 °C
Salinity Correction 0.0 to 50.0 ppt,
Barometric Pressure Correction 555 to 1499 mm Hg or 66.6 to 199.9 kPA,
Method Automatic correction after manual input
No. of Calibration Points Up to 2 points; 100% in air and/or 0% in known solution
Special Functions Self-diagnostic
Electrode characteristics Hold & Auto-Power Off after 20 minutes



To locate your nearest distributor click here

Please click here to the previous page.

DISTRIBUTORS

Click here to view world map
CONTACT US
WPA
Unit 22 - Cambridge Science Park
Milton Road
Cambridge
CB4 0FJ

Tel:
+44 (0) 1223 423723
Fax:
+44 (0) 1223 420164

E-mail:
To contact WPA please click here
COPYRIGHT © WPA - MOST RECENT REVISION 21.11.2008



Website designed by mi